    Brand Name : OEM
    Certification : CE / FDA
    Price : Negotiable
    Delivery Time : 3 working days
    Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram
    • Product Details
    • Company Profile

    Portable Medical Ventilator Machine , Ventilator Oxygen Machine For First Aid Room

    Portable Medical Ventilator Machine , Ventilator Oxygen Machine For First Aid Room


    Technical Specifications:

    Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa

    safety valve <12.5 kPa

    Respiratory Rate 1~99 bpm

    gas concentration output 21% ~ 100% adjustable

    Maximum Minute ventilation ≥18L / min, permissible deviation: ±20 %

    Breathing Mode A/C, SIPPV,IPPV,SIMV,PSIMV,CPAP,PCV, PEEP,SIGN,VCV, SPONT, MANUAL

    Ventilator compliance ≤30 Ml/kPa

    PEEP: 0~1.0KPa

    Ptr: ,-1.0~1.0 kPa

    IMV frequency: 1 ~ 12 times/min, permissible deviation: ±15%.

    I/E Ratio: 4:1~1:4

    Tidal Volume range 0~2000ml

    Inspiratory Plateau: 0~1s

    Maximum Safety Pressure ≤ 6.0kPa

    Sigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point and the permissible error is -10% or 15%.

    Pressure limit range: 1.0~ 6.0kPa

    Airway pressure alarm: Audible and visual and with yellow and red color indicating upper alarm: 0.9 kPa ~ 5.4 kPa

    lower: 0.5kPa

    Ventilation Volume alarm: ≥ 18L/ MIN

    Power Supply Alarm "Ac/dc power supply are after failing to send out alarm immediately

    Alarm Time: keep >120s"

    The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.


    Ambient temperature10 ~ 40oC
    Relative humidityno higher than 80 %
    Atmospheric pressure860 hPa ~ 1060 hPa
    Power requirement100-120 Vac, 50/60 Hz;
    Air source requirementMedical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa.
    Attentionthe AC power supply used for the anesthesia machine must be well grounded.
    Attentionthe anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment
    Part IISpecial Requirements for Safety and Basic Performance of Anesthesia System.